EasyManuals Logo

Mindray A5 Service Manual

Mindray A5
436 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #246 background imageLoading...
Page #246 background image
System Calibration Calibration
4 - 60 046-001141-00 A5/A3â„¢ Service Manual
4.3.5.2 Troubleshoot Pressure and Flow Zeroing Failure
In case of zeroing failure, troubleshoot as follows:
1. Set the anesthesia machine to manual ventilation or standby mode. Turn off fresh gas. Unplug
the breathing tubes in the breathing system, causing the inspiration and expiration connectors
to open to the air. Bleed the residual gas inside the bellows. Make sure that there is no flow or
pressure entering the flow or pressure sensors inside the machine.
2. Check if the zero points of the sensors are normal by referring to 3.7.3 Check the Sensor Zero
Point.
3. If a zero point error is detected, unplug the sensor sampling line to eliminate the effects caused
by sampling line occlusion or three-way valve. If zero point is still out of the range, the ventilator
control board is faulty. Replace the ventilator control board.
4. If zero points of the sensors are correct but zeroing is still failed, the solenoid valve assembly is
faulty. Replace the solenoid valve assembly.
4.3.6 Electronic Flowmeter Zeroing (user)
After the gas supply is disconnected, if the pointer of the pressure gauge returns to zero but the
electronic flowmeter still displays flow, it is possible that zero offset occurs to the electronic
flowmeter's sensor. Generally, you can zero the flowmeter manually to eliminate the measurement
deviation caused by zero offset immediately.

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
TypeAnesthesia Machine
BrandMindray
ModelA5
Power Supply100-240 V AC, 50/60 Hz
Ventilation ModesSIMV, PSV
FlowmetersO2, N2O, Air
Vaporizer CapacityUp to 3 vaporizers
Safety FeaturesO2 failure protection, low pressure alarm
Gas ModuleOptional integrated gas module
Integrated Air CompressorOptional
Patient MonitoringIntegrated monitoring with optional modules

Related product manuals