2.1 Preparation - Additional Material Required
The following additional material are required before installation. The customer is responsible for
supplying this material. Missing items may result in delays, incomplete installations, and/or additional
service visits.
• Compatible emergency O2, N2O, and AIR cylinders
• Agent vaporizers and key fillers (if not purchased with the anesthesia system)
• Liquid agent medication
• CO2 absorbent Pre-Paks or loose fill
• Active O2, N2O, and AIR lines (280 to 600 kPa (40 to 87 psi))
• Dropdown hoses for ceiling-mounted medical gas utilities that are compatible with quick-
disconnect hoses (if not purchased with the anesthesia system)