Functional Tests Installation Guide
2 - 34 046-001141-00 A5/A3™ Service Manual
FIGURE 2-49 100% O2 Sensor Calibration Canceled
7. The screen shown below is displayed if the calibration has failed. A Fail code is displayed in red.
Select Try Again to repeat the calibration. Select Done to exit the calibration screen.
FIGURE 2-50 100% O2 Sensor Calibration Failed
8. The screen as shown below is displayed after a successful calibration. Select Done to exit the
calibration screen.