System Calibration Calibration
4 - 16 046-001141-00 A5/A3â„¢ Service Manual
Follow these steps to calibrate flow sensors.
1. Enter Standby.
2. Select Setup-> Calibrate-> Calibrate Flow Sensor (version 1.03.02 and earlier) or Setup->
General-> Calibrate Flow Sensor (version 2.01.00 and higher) to access the screen shown below.
FIGURE 4-11
3. Set up the machine as per the instructions on the screen and select Next to open the menu
shown below.
FIGURE 4-12
4. Select Begin to calibrate flow sensors. During the calibration, you can select Cancel to cancel the
calibration.