A5/A3™ Service Manual 046-001141-00 4 - 27
Calibration System Calibration
FIGURE 4-32 Pneumatic connection with the calibration device (VT Plus)
14. Press the Continue button to open the menu shown below.
FIGURE 4-33
15. The screen shown below is displayed after the calibration is completed.
The screen shown below is displayed if the flow sensor calibration has failed. When the
calibration has failed, read the screen of the calibration device for further information on the
cause of the failure. Select Try Again to do the calibration again. Select Done to exit the
calibration screen.