System Calibration Calibration
4 - 32 046-001141-00 A5/A3™ Service Manual
FIGURE 4-40
6. The menu shown below is displayed after the first step of manual calibration is completed.
FIGURE 4-41
7. Calibration Device Setup (TSI Certifier 4070):
a. Flow Setting: Press the to line select key until the top display reads LPM or SLPM. If the
display reads SLPM, press the DISPLAY UNITS key until the top display reads LPM.
b. Gas Settings: Press the GAS SELECT key until the bottom left corner reads O2.