System Calibration Calibration
4 - 34 046-001141-00 A5/A3™ Service Manual
FIGURE 4-44 Pneumatic connection with the calibration device (TSI Certifier 4070)
11.Press the Accept button and the screen shown below is displayed.
The screen shown below is displayed after a successful flow sensor calibration.
FIGURE 4-45
The screen shown below is displayed if the flow sensor calibration has failed.