Technical Alarms Check Repair and Troubleshooting
5 - 14 046-001141-00 A5/A3â„¢ Service Manual
Iso Accuracy Unspecified Low
The measured value
exceeded the module
accuracy range.
1. Re-calibrate the AG module.
2. Replace the AG module.
Sev Accuracy Unspecified Low
The measured value
exceeded the module
accuracy range.
1. Re-calibrate the AG module.
2. Replace the AG module.
Hal Accuracy Unspecified Low
The measured value
exceeded the module
accuracy range.
1. Re-calibrate the AG module.
2. Replace the AG module.
Des Accuracy Unspecified Low
The measured value
exceeded the module
accuracy range.
1. Re-calibrate the AG module.
2. Replace the AG module.
Mixed Agent Low
The AG module
supported monitoring
and calculation of two
kinds of halogenated
anesthetic agents, and
the measured MAC was
less than 3.
Use only one halogenated anesthetic agent.
Mixed Agent Med
The AG module
supported monitoring
and calculation of two
kinds of halogenated
anesthetic agents, and
the measured MAC was
less than 3. And the
monitored value of one
agent was invalid.
Use only one halogenated anesthetic agent.
Mixed Agent and MAC≥3 Med
The AG module
supported monitoring
and calculation of two
kinds of halogenated
anesthetic agents, and
the measured MAC was
greater than or equal to
3.
Use only one halogenated anesthetic agent.
External AG Module
Disconnected
High
FDA version had external
AG module in standard
configuration, but the
external AG module was
not connected.
Load the external AG module.
Incompatible AG Software
Version
High
AG module software
version was lower than
1.7.3.
1.Replace with AG module of high version software.
2.Return to the factory to update the low-version AG
module of low version.
EtCO2 Overrange Low
The monitored value
exceeded the module
measurement range.
1. Reduce the concentration of the monitored gas to the
normal range.
2. Re-calibrate the AG module.
3. Replace the AG module.
FiCO2 Overrange Low
The monitored value
exceeded the module
measurement range.
1. Reduce the concentration of the monitored gas to the
normal range.
2. Re-calibrate the AG module.
3. Replace the AG module.

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
BrandMindray
ModelA5
CategoryMedical Equipment
LanguageEnglish

Related product manuals