Technical Alarms Check Repair and Troubleshooting
5 - 16 046-001141-00 A5/A3â„¢ Service Manual
EtDES Overrange
Low
Low
The monitored value
exceeded the module
measurement range.
1. Reduce the concentration of the monitored gas to the
normal range.
2. Re-calibrate the AG module.
3. Replace the AG module.
FiDES Overrange
Low
Low
The monitored value
exceeded the module
measurement range.
1. Reduce the concentration of the monitored gas to the
normal range.
2. Re-calibrate the AG module.
3. Replace the AG module.
Rate Overrange Low
The monitored value
exceeded the module
measurement range.
1. Reduce the concentration of the monitored gas to the
normal range.
2. Re-calibrate the AG module.
3. Replace the AG module.