Page #340 background imageLoading...
Page #340 background image
Software Update and Software Configuration Activation Repair and Troubleshooting
5 - 78 046-001141-00 A5/A3â„¢ Service Manual
NOTE: It is normal for the power board software that it will fail at the first time
and it will be successful at the second time in upgrade process.
15.Once the last file is upgraded turn off the A5/A3 unit and then restart A5/A3 unit.
16.Check the software version on the A5/A3 by clicking Setup then Service, enter the service
password "789789" followed by Enter, then go to System Info and then SW Versions. Verify
that the software version on the A5/A3 match the following table.
NOTE: This only an example of a Software version table. Check for Technical
bulletins to find the correct table for the software version you are
installing.
Software version V01.02.00
Module Software Version Date*
Host Software 01.02.00 1/19/2011
BIOS 01.03.00.00 11/29/2011
FPGA Display 1.2 \
FPGA Sound 1.0 \
Ventilator Protect Module V1.1 11/26/2011
Ventilator Control Module V1.3 3/04/2011
Power System V1.3 1/19/2011
Flowmeter Software V1.2 1/19/2011
* The Date format my differ depending on the unit setup

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
BrandMindray
ModelA5
CategoryMedical Equipment
LanguageEnglish

Related product manuals