Disassemble the Assemblies Repair and Disassembly
6 - 36 046-001141-00 A5/A3â„¢ Service Manual
3. Remove the 11 screws around the display mount then remove the display front cover.
FIGURE 6-65
4. Slightly lift the touch screen off the display front cover to remove the touch screen.
FIGURE 6-66
6.2.4.7 Remove the Display Replacement Package
1. Remove the display interface board.
2. Remove the touch screen control board.
3. Remove the touch screen.