BeneVision N Series Patient Monitor Operator’s Manual C - 11
Art-S Extreme Alarm switch Off
High limit Adult: 175 mmHg
Pediatric: 130 mmHg
Neonate: 95 mmHg
Low limit Adult: 75 mmHg
Pediatric: 60 mmHg
Neonate: 50 mmHg
Priority High
Alarm Outputs Off
Art-D Extreme Alarm switch Off
High limit Adult: 105 mmHg
Pediatric: 80 mmHg
Neonate: 65 mmHg
Low limit Adult: 35mmHg
Pediatric: 30 mmHg
Neonate: 15 mmHg
Priority High
Alarm Outputs Off
Art-M Extreme Alarm switch Off
High limit Adult: 125 mmHg
Pediatric: 100 mmHg
Neonate: 75 mmHg
Low limit Adult: 55 mmHg
Pediatric: 40 mmHg
Neonate: 30 mmHg
Priority High
Alarm Outputs Off
CPP Alarm switch On
High limit Adult: 130 mmHg
Pediatric: 100 mmHg
Neonate: 90 mmHg
Low limit Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 30 mmHg
Priority Med
Alarm Outputs Off
Measure (for P1, P2) All
Measure (for P3, P4) Mean only
Sensitivity Med
Speed 25 mm/sec
Item Default Setting