Monitoring NIBP Configuring NIBP Measurement Modes
BP10 Operator’s Manual 6 - 7
6.6 Configuring NIBP Measurement Modes
BP10 provides five measurement modes:
Manual mode
Continuous mode
Automatic mode
Sequence mode
ABPM (Ambulatory Blood Pressure Monitoring) mode
The automatic, sequence and ABPM modes need to configure more settings.
NOTE
You can initiate a manual or continuous measurement in the course of
automatic, sequence and ABPM series measurements.
6.6.1 Setting the Automatic Mode
In automatic mode, the NIBP measurements are automatically taken on the configured
interval.
To enable and set up the automatic mode, follow this procedure:
1. In the main menu, select Mode and set it to Auto.
2. Select Interval and select the appropriate period. The default setting is 15 min.
6.6.2 Setting the Sequence Mode
In the sequence mode, the NIBP measurements are automatically taken in the user-
defined sequence, which allows you to set different measurement intervals in
consecutive five (5) phases.
To enable and set up the sequence mode, follow this procedure:
1. In the main menu, select Mode and set it to Sequence.
2. Select Sequence Settings and open the Sequence Settings screen.
3. Set up the interval and duration in each phase.
Options Description Settings
Phase In a sequence, you can set up to five
measurement phases from P1 to P5.
Note: If some existing phases need
to be removed, set the phase
interval to OFF.
/