Product Specifications Power Supply Specifications
BP10 Operator’s Manual A - 3
A.3 Power Supply Specifications
A.3.1 BP10
BP10 is powered by batteries.
A.3.2 Central Charger
The central charger is powered by external AC power.
Battery type
AA batteries (two)
Rechargeable lithium-ion battery (one)
Capacity for lithium-ion
battery
1800 mAh
Charge time
At the room temperature: ≤ 5 hours
The charge time definition: the time for the battery from the
exhausted status to 90% battery charge (within the operating
temperature range declared for the unit).
Safety and authentication
for lithium-ion battery
UL2054, IEC62133, UN38.3
Run time
In the condition of turning off the display, the batteries perform
as follows.
One lithium-ion battery: At least 700 NIBP measurements
Two AA batteries: At least 300 NIBP measurements
Shutdown delay in low
battery
At least 10 NIBP measurements after the Low Battery message
first occurs.
At least 30 seconds after the Battery Depleted message first
occurs.
Input voltage
100 VAC to 240 VAC (± 10%)
Frequency
50 Hz/60 Hz (±3 Hz)
Input current
1.5 A to 0.75 A
Overcharge protection
function
The charger automatically stop charging when the lithium-ion
battery charge is full.