NIBP Measurement Specifications Product Specifications
A - 6 BP10 Operator’s Manual
A.7 NIBP Measurement Specifications
Standard
Meet standards of IEC80601-2-30
Te ch ni qu e
Oscillometry
Mode of operation
Manual, Automatic, Continuous, Sequence and ABPM
Automatic mode
repetition intervals
1min, 2min, 2.5min, 3min, 5min, 10min, 15min, 20min, 30min, 1h,
1.5h, 2h, 3h, 4h or 8h.
Maximum measurement
time
Adult and pediatric: 120 seconds
Measurement ranges
(mmHg)
Systolic Diastolic Mean
Adult 25 to 290 10 to 250 15 to 260
Pediatric 25 to 240 10 to 200 15 to 215
Accuracy
Maximum mean error: +/- 5 mmHg
Maximum standard deviation: 8 mmHg
Resolution
1 mmHg
Static pressure
measurement range
0 to 300 mmHg
Static pressure
measurement accuracy
+/-3 mmHg
Initial cuff inflation
pressure (mmHg)
Range Default Step
Adult 80 to 280 160 10
Pediatric 80 to 210 140 10
Auxiliary Venipuncture
(mmHg)
Range Default Step
Adult 20 to 120 80 10
Pediatric 20 to 80 60 10
Pulse Rate
Measurement range
30 bpm to 300 bpm
Accuracy
+/-3 bpm or +/-3%, which ever is greater
Resolution (bpm)
1 bpm