EasyManua.ls Logo

Mindray BP10 NIBP Module - Setting up the ABPM Mode; About Continuous Mode

Mindray BP10 NIBP Module
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuring NIBP Measurement Modes Monitoring NIBP
6 - 8 BP10 Operator’s Manual
6.6.3 Setting up the ABPM Mode
In ABPM mode, the NIBP measurements are automatically taken according to the
configured intervals for the day and the night.
To enable and set up the ABPM mode, follow this procedure:
1. In the Menu screen, select Mode and set it to ABPM.
2. Select ABPM Settings and open the ABPM Settings screen.
Set the start time and interval for the day and night respectively.
The default start time of the day is 6:00, and the interval is 15 min.
The default start time of the night is 22:00, and the interval is 30 min.
6.6.4 About Continuous Mode
In the continuous mode, the measurements are taken one after another in a five-minute
period. After that, the continuous mode ends, and the device restores to the NIBP mode
before the continuous mode starts.
To enable the continuous mode, enter the main menu and select NIBP STAT.
Interval Set up the interval between NIBP
measurements in an individual phase.
1 min, 2 min, 2.5 min, 3 min,
5 min, 10 min, 15 min, 20
min, 30 min, 1 h, 1.5 h, 2 h,
Off.
Duration Set up the lasting time of an individual
phase.
10 min, 30 min, 1 h, 2 h, 4 h,
8 h, continuous
Options Description Settings

Table of Contents