Charging the Rechargeable Lithium-ion Battery Battery
8 - 4 BP10 Operator’s Manual
Remove the battery before transporting the device or if the device is not in
use or is being stored.
NOTE
Retain the NIBP hose with the device while removing the battery.
1. Lift up the lithium-ion battery pack or AA alkaline battery tray at the bottom of
BP10.
2. Remove the lithium-ion battery pack or AA alkaline battery tray from BP10.
8.6 Charging the Rechargeable Lithium-ion Battery
WARNING
Only use the specified central charger to charge to the lithium-ion batteries
designated by Mindray.
Only use the approved power cord with the grounded mains plug to firmly
connect the central charger to a grounded AC mains socket. Never refit the
mains plug to fit an ungrounded AC mains socket.
Do not use the Multiple Portable Socket Outlets (MPSO) or AC mains
extension cords. Use an IEC 60601-1 approved isolation / separation
transformer, otherwise, it may result in leakage current. Insure that the sum
of the individual ground leakage currents does not exceed the allowable
limits.
Do not place any shield object (such as cloth or paper) to cover the central
charger or batteries, and keep ventilated while charging the lithium-ion
batteries.
Do not connect other devices to the power supply system.
Do not use the central charger to charge the lithium-ion batteries in high
temperature above 40°C.