General Problems Troubleshooting
12 - 2 BP10 Operator’s Manual
12.1 General Problems
12.2 Technical Messages and Solutions
The following table lists the major technical messages displayed in the message area of
BP10.
For the clearable technical messages, you can clear them by selecting
Clear Prompt in the
main menu.
Symptom Possible Cause Solution
Cannot discharge patient, change
patient category, or set NIBP
measurement related configurations
at BP10.
BP10 is paired with TD60/TM80. Unpair BP10 from TD60/TM80 by
pressing the key, and then
discharge the patient by
selecting Discharge Patient in
the main menu at BP10.
The BP10 cannot be paired with the
TD60/TM80.
1. The distance between BP10
and TD60/TM80 is out of the
radio range.
2. Signal is interrupted.
3. The TD60/TM80 is already
paired with other BP10.
4. After BP10 is paired with TD60/
TM80, user turns off and then
turns on the MPAN at BP10 or
TD60/TM80.
Restart the pairing atTD60/
TM80.
Message
Clearable
(Yes/No)?
Possible cause Solution
Low Battery No The battery charge is low. Replace with a new battery.
Battery Depleted No The battery charge is almost
depleted.
Voltage Error No The lithium-ion battery voltage is
out of the specified range.
Battery Error No The lithium-ion battery
communication is error.
Verify that the battery is
correctly installed.
If the error persists, replace with
a new battery.
Battery Maintenance
Required
No The lithium-ion battery is aging. Replace with a new battery.
MPAN Disconnected Yes 1. The distance between BP10
and TD60/TM80 is out of the
radio range.
2. The TD60/TM80 is powered off.
1. Move the BP10 and TD60/
TM80 closely.
2. Power on the TD60/TM80.