Troubleshooting Technical Messages and Solutions
BP10 Operator’s Manual 12 - 3
Clock Needs to Be Reset Yes The button cell does not have
sufficient charge.
Reset the system time.
NIBP Module Error No ■ An error occurred to the
NIBP module.
■ There is a problem with
communications between
the module and the
telemetry.
Restart the device.
NIBP Cuff Loose Yes ■ The NIBP cuff is not
properly connected.
■ There is a leak in the
airway.
1. Check the patient’s
condition and verify
patient type.
2. Replace with an appropri-
ate cuff and connect it
correctly.
NIBP Airway Error Yes The device cannot finish one
correct zeroing.
Restart the device.
If the error persists, contact your
service personnel.
NIBP Weak Signal Yes The patient’s pulse is weak or the
cuff is loose.
Check the patient condition and
change the cuff application site.
If the error persists, replace the
cuff.
NIBP Overrange Yes The measured NIBP value is not
within the measurement range.
Contact your service personnel.
NIBP Excessive Motion Yes Patient’s arm moves too much. Check the patient’s condition
and reduce the patient motion.
NIBP Cuff Overpressure Yes The NIBP airway may be
occluded.
Check the airway and measure
again.
NIBP Cuff or Airway
Leak
Yes The NIBP airway may leak air. 1. Verify that the cuff is
properly connected.
2. Verify that the airway
does not leak air.
NIBP Timeout Yes ■ Time is out.
■ The measurement time is
over 120 seconds.
1. Check the patient’s
condition and NIBP
connections.
2. Replace the cuff.
NIBP Cuff and Patient
Mismatch
Yes The cuff type applied mismatches
the patient category.
Verify the patient category and
replace the cuff.
Intervals Not Set Yes No interval in Sequence mode is
not set.
Set the intervals.
Message
Clearable
(Yes/No)?
Possible cause Solution