Classifications Product Specifications
A - 2 BP10 Operator’s Manual
A.1 Classifications
The Telemetry Monitoring System is classified, according to IEC60601-1:
A.2 Environmental Specifications
• The device may not meet the performance specifications if stored or used
outside the specified temperature and humidity ranges.
Type of protection against
electrical shock
BP10: energized from an internal electrical power source.
Central Charger: Class I.
Degree of protection against
electrical shock for BP10
Type CF defibrillation proof for NIBP.
Mode of operation
Continuous
Degree of protection against
harmful ingress of water
BP10: IP32
Central Charger: IPX0
Degree of protection against
hazards of explosion
Not suitable: Equipment not suitable for use in the presence
of a flammable anesthetic mixture with air with oxygen or
nitrous oxide.
Degree of protection against
hazard of dropping for BP10
No damage by dropping from a height of 1.5m.
Sterilization or disinfection
According to the method(s) of sterilization or disinfection
recommended by the manufacturer.
Item Operating conditions Storage conditions
Temperature 0 °C to 40 °C -20 °C to 60 °C
Relative humidity
(noncondensing)
15% to 95% 10% to 95%
Barometric 427.5 mmHg to 805.5 mmHg, or
57.0 kPa to 107.4 kPa
120 mmHg to 805.5 mmHg, or
16.0 kPa to 107.4 kPa