Requirements of Performance Indices C-1
Appendix C Requirements of
Performance Indices
Probe
model
Freq.
Lateral
resolution
(mm)
Axial
resolution
(mm)
Blin
d
area
(mm)
Detecti
on
depth
(mm)
Geometric
positioning
accuracy %
35C50P
3.5
≤2 ( depth≤80)
≤4 (80<depth
≤130)
≤1 (depth≤
80)
≤2 (80<
depth≤130)
≤3
≥170
transverse≤5
lateral≤3
75L38P
7.5
≤1 (depth≤60)
≤1 (depth≤
80)
≤2
≥80
transverse≤4
lateral≤3
6CV1P
6.5
≤1(depth≤40)
≤1(depth≤
60)
≤3
≥80
transverse≤3
longitudinal≤
3
2P2P
2.5
≤3(depth≤80)
≤1(depth≤
80)
≤3
≥180
transverse≤3
longitudinal≤
3
≤4(80<depth≤
160)
≤2 (80<
depth≤170)
7L5P
7.5
≤1(depth≤60)
≤0.5(depth
≤80)
≤2
≥80
transverse≤3
longitudinal≤
3
6C2P
6.5
≤1 (depth≤40)
≤1 (depth≤
60)
≤3
≥80
transverse≤3
longitudinal≤
3
7L4P
7.5
≤1(depth≤60)
≤0.5(depth
≤80)
≤2
≥80
transverse≤3
longitudinal≤
3
CB10-4P
6.5
≤1(depth≤40)
≤1(depth≤
60)
≤3
≥80
transverse≤3
longitudinal≤
3
7LT4P
7.5
≤1 (depth≤60)
≤1 (depth≤
80)
≤2
≥80
transverse≤3
lateral≤3
3C5P
3.5
≤1(depth≤80)
≤1(depth≤
80)
≤3
≥180
transverse≤3
longitudinal≤
3
≤2(80<depth≤
130)
≤2(80<depth
≤130)
V11-3P
6.5
≤1(depth≤40)
≤1(depth≤
60)
≤3
≥80
transverse≤3
lateral≤3
P4-2P
3.0
≤3(depth≤80)
≤1(depth≤
80)
≤3
≥180
transverse≤3
lateral≤3