Software installation and Maintenance 6-1
6 Software installation and
Maintenance
WARNING: DO NOT directly remove a USB memory device; otherwise,
the USB memory device and / or the system may be damaged.
6.1 Enter the Maintenance Window
NOTE:
Log on the system with the identity of Service before perform system
maintenance.
To log on the system:
1. When access control function has not been activated: press “Ctrl+/” to show the
Login dialogue box, and then select the Service as the user name.
2. When access control function has been activated already: press “Ctrl+/” when the
Login dialogue box is displayed, and then select the Service as the user name.
3. Press <Setup> key to open the Setup menu, click [Maintenance] to enter [Other]
menu and then select the target items to perform the maintenance respectively.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-30 and is the answer not in the manual?

Mindray DC-30 Specifications

General IconGeneral
DICOMYes
Hard Drive500 GB
Continuous Wave DopplerYes
Tissue Harmonic ImagingYes
Relative Humidity30% to 85% (non-condensing)
Imaging ModesColor Doppler, Power Doppler
Power Supply100-240 V~ 50/60 Hz
Probe Connectors2 active probe connectors
Probe TypesConvex, Linear, Phased Array
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +55°C
Transducer Ports2