Function Checking and Testing 5-15
5.5.2.3 Geometric positioning accuracy
Longitudinal geometric positioning accuracy
Test Step:
1. Do adjustments as the way in testing the maximum depth.
2. Record the distance by 20mm each segment on the longitudinal targets line using
the measurement caliper;
3. Select the value with the greatest error (to 20mm), calculate the accuracy using the
formula below