6-4 Software installation and Maintenance
[Start] to log in the system. The system is no longer limited to the
installment.
e) Global password: used to protect the installment from being changed.
CAUTION:
The global password is displayed only once, please
remember the global password carefully. If you forget the
global password, you cannot log on the system with the
identity of service and reset the installment after setting
the installment and restarting the system.
3. Click [Reset] to reset the installment. Set the times of the installment in “Periods” list
and the time interval of each installment in “Days” list. Then, click [Generate], and
tick off [Display Password]. The new installment password generated by the system
is displayed.
4. Insert USB device. Click [Save As] to select the directory, and then click [OK] to save
the installment password, as shown below.
5. Restart the system. Click [Service entry], input the Global Password, click [Login],
and the Service account login dialog is displayed. Input the service account
password, click [Login], and the system enters the Doppler interface.
6. Press [Setup]. Select [Maintenance] -> [Other] -> [Installment]. Click [Set
Installment], a login dialog is displayed. Input the global password to enter the Set
Installment interface.
View the installment password
Press [Setup]. Select [Maintenance] -> [Other] -> [Installment], and tick off [Display
Password] to view the installment password. Or
Open txt. file which contains the password, and view the installment password.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-30 and is the answer not in the manual?

Mindray DC-30 Specifications

General IconGeneral
DICOMYes
Hard Drive500 GB
Continuous Wave DopplerYes
Tissue Harmonic ImagingYes
Relative Humidity30% to 85% (non-condensing)
Imaging ModesColor Doppler, Power Doppler
Power Supply100-240 V~ 50/60 Hz
Probe Connectors2 active probe connectors
Probe TypesConvex, Linear, Phased Array
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +55°C
Transducer Ports2

Related product manuals