EasyManua.ls Logo

Mindray Eagus TEX21 - Page 257

Mindray Eagus TEX21
370 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13 Probes and Biopsy
Operator’s Manual 13 - 53
b. Turn on the knob of fixing guiding block and remove the guiding block from the hole of
installing guiding block.
c. Turn on the knob of fixing needle-guided bracket and remove the needle-guided bracket.
NGB-058
Perform the following procedure:
1. Install the needle-guided bracket:
a. Put on the sterile probe sheath.
b. Hold the probe in one hand, select the correct needle-guided bracket and hold it with the
other hand. Match the inside of the bracket with the outside of the probe respectively.
Mount the bracket onto the probe. Rotate the locking nuts of the guided bracket at the
right and left sides to fix the bracket and the probe. Rotate the needle-shaped adjusting nut
to the ultimate position as shown in the figure.
1 Right clamp of the
bracket
2 Left clamp of the bracket 3 Angle locking nut
4 Angle adjusting block 5 Needle clamping cover 6 Angle adjusting base
7 Biopsy needle pressure
position
8 Torsion spring of the
needle pressure cover
9 Locking nuts of the
guided bracket at the
right and left sides
<2><1> <4> <5>
<6>
<7><8><9>
<3>

Table of Contents