EasyManuals Logo

Mindray TE Air User Manual

Mindray TE Air
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
Revision:1.0(2023–5–17)
44
2.
In charging state, the three-segment battery indicator and the lightning flash status indicator on
the wireless probe act as follows:
1)When the battery level is lower than 20%, the battery indicator is off, and the status indicator
stays orange.
2)When the battery level is equal to or higher than 20% and lower than 40%, segment 1 of the
battery indicator blinks, and the status indicator stays white.
3)When the battery level is equal to or higher than 40% and lower than 70%, segment 1 of the
battery indicator stays on, segment 2 blinks, and the status indicator stays white.
4)When the battery level is equal to or higher than 70% and lower than 95%, segments 1 and 2
of the battery indicator stay on, segment 3 blinks, and the status indicator stays white.
5)When the battery level is equal to or higher than 95%, all segments of the battery indicator stay
on, and the status indicator stays white.
When the both battery indicator and status indicator stay white, the wireless probe has been fully
charged.
Diagnostic Ultrasound System
Service Manual
3 How to Use

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TE Air and is the answer not in the manual?

Mindray TE Air Specifications

General IconGeneral
BrandMindray
ModelTE Air
CategoryMedical Equipment
LanguageEnglish

Related product manuals