Revision:1.0(2023–5–17)
27
• Charging status indicator: When the wireless probe is not in the charging station, the indicator
is off. When the probe is being charged in the charging station, the indicator stays orange.
When the probe is fully charged, the indicator stays green.
Wireless Probe Charging Case (Air Capsule)
Figure2–5 Charging Case (Air Capsule) and charging mechanism
The Charging Case (Air Capsule) has a 6000 mAh battery and supports charging via a Type-C
cable and wireless charging via the Qi protocol.
Main function:
• When the Charging Case (Air Capsule) detects presence of the probe, it provides 5V output to
charge the probe.
• During the charging process, the Charging Case (Air Capsule) measures the battery level of
the probe.
• When the probe is in the Charging Case (Air Capsule), the Charging Case (Air Capsule)
charges the probe and starts to charge its battery only after the probe is fully charged.
• When the Charging Case (Air Capsule) is connected to a Type-C charger and a wireless
charger at the same time, only the Type-C charger works.
• The indicator on the Charging Case (Air Capsule) stays orange when the battery level of the
Charging Case (Air Capsule) is 0%-30%, blue when the battery level is 30%-60%, and white
when the battery level is 60%-100%.
22..33..33..22 Battery Management Module
The battery management module provides the following functions:
1.
Control of probe power-on/off
(Description of the power-on time process)
2. Battery charging and discharging management
3. Control of the power key and battery indicator
Diagnostic Ultrasound System
Service Manual
2 Product Knowledge