Revision:1.0(2023–5–17)
53
5 Maintenance
5.1 Overview
5.1.1 Description
This chapter describes only the recommended maintenance processes.
5.1.2 Tools, Measurement Devices and Consumables
Table 5–1
Tools and measuring devices
Tool or Measuring Device Quantity Remarks
Plastic or resin container 1 Used to store the saline, and
can accommodate two
ultrasound probes.
Soft brush 1 Its size is similar to that of a
toothbrush, and it has a soft
head.
Small plastic basin 1 Used to store the soapy
water.
Safety tester 1 Used when conducting
electrical safety tests. For
details, see
8.2
Electrical
Safety Maintenance
Table 5–2 Tools and measuring devices
Consumables Quantity Remarks
Aluminum foil About 1 m None
Normal saline solution About 1000 mL At least half of the container
should be filled with the
solution so that the probe can
be submerged by the solution.
(Concentration: 0.85-0.95%)
Mild soapy water About 400 mL None
Dry soft cloth or cotton cloth About 5 pcs None
Diagnostic Ultrasound System
Service Manual
5 Maintenance