Revision:1.0(2023–5–17)
73
2.
The Charging Case (Air Capsule) is faulty.
3. The probe is faulty.
Troubleshooting
1. Keep the probe still for its temperature to drop, and then check whether it can be charged.
2. Check whether the probe can be charged using the charging cable.
3. If the probe can be charged using the charging cable, the Charging Case (Air Capsule) is
faulty. Otherwise, the probe is faulty.
Solution
1. Wait until the probe temperature drops, and then charge it again.
2. Replace the Charging Case (Air Capsule).
3. Replace the probe. For details, see the instructions for the i3P ultrasound probe (FDA) 6.1.2
Disassembly and Assembly
.
77..22..66..22 Slow Probe Charging
Fault Description
The probe is charged slowly.
Involved FRU
Wireless probe and its Charging Case (Air Capsule)
Severity
Maintenance as soon as possible
Response
Contact the device provider for onsite troubleshooting.
Possible Causes
1. The charging port is not reliably connected.
2. The Charging Case (Air Capsule) is faulty.
3. The ambient temperature is high.
4. The probe is faulty.
Troubleshooting
1. Remove and reconnect the probe and check whether the charging speed is improved.
Diagnostic Ultrasound System
Service Manual
7 Troubleshooting