Mindray TE Air - Page 74

Mindray TE Air
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Revision:1.0(2023–5–17)
74
2.
Check whether the probe can be charged normally using the charging cable.
3. Charge the probe at the room temperature and check whether the charging speed is improved.
4. If the probe can be charged normally using the charging cable, the Charging Case (Air
Capsule) is faulty. Otherwise, the probe is faulty.
Solution
1. Remove and reconnect the probe properly.
2. Replace the Charging Case (Air Capsule).
3. Replace the probe. For details, see the instructions for the i3P ultrasound probe (FDA) 6.1.2
Disassembly and Assembly
.
Diagnostic Ultrasound System
Service Manual
7 Troubleshooting