1.
Replace the probe. For details, see the instructions for the i3P ultrasound probe (FDA) 6.1.2
Disassembly and Assembly
.
2. Upgrade the software to fix the bugs.
7.2.2 Connection Failures
77..22..22..11 No Wi-Fi Hotspot
Fault Description
The Wi-Fi hotspot is not found.
Involved FRU
Wireless probe
Severity
Suspension
Response
Contact the device provider for onsite troubleshooting.
Possible Causes
The probe is faulty.
Troubleshooting
Check that the probe is turned on, and use two mobile phones to search for the Wi-Fi hotspot. If
the Wi-Fi hotspot is still not found, the probe is faulty.
Solution
1. Reset the probe and try again.
2. Replace the probe. For details, see the instructions for the i3P ultrasound probe (FDA) 6.1.2
Disassembly and Assembly
.
77..22..22..22 Low Signal Strength of the Wi-Fi Hotspot
Fault Description
The signal strength of the Wi-Fi hotspot is low.