The probe cannot be charged.
Involved FRU
Wireless probe and its charging station
Severity
Maintenance as soon as possible
Response
Contact the device provider for onsite troubleshooting.
Possible Causes
1. The temperature inside the probe is high.
2. The charging station is faulty.
3. The probe is faulty.
Troubleshooting
1. Keep the probe still for its temperature to drop, and then check whether it can be charged.
2. Check whether the probe can be charged using the charging cable.
3. If the probe can be charged using the charging cable, the charging station is faulty. Otherwise,
the probe is faulty.
Solution
1. Wait until the probe temperature drops, and then charge it again.
2. Replace the charging station.
3. Replace the probe. For details, see the instructions for the i3P ultrasound probe (FDA) 6.1.2
Disassembly and Assembly
.
77..22..44..22 Slow Probe Charging
Fault Description
The probe is charged slowly.
Involved FRU
Wireless probe and its charging station
Severity
Diagnostic Ultrasound System
Service Manual
7 Troubleshooting