Revision:1.0(2023–5–17)
54
5.1.3 Routine Maintenance Items
Table 5–3
Maintenance items and maintenance frequency
SN
Item Recommend-
ed
Frequency
Operator
Method
End user Engineer
1 Cleaning the
probe
After each use
See the
cleaning
procedure.
2 Checking the
probe surface
1 time/day
For
details,
see
5.3.1
General
Inspections
3 Checking the
charging
cable and
adapter
Once per
month
4 Checking the
battery
Once per 3–6
months
5
Checking the
functions of
peripherals
and optional
accessories
Once per year
6 Electrical
safety
inspection
Once per year
5.2 System Cleaning
5.2.1 Cleaning the Probe
1. Wear gloves to prevent contamination.
2. Turn off the probe. If the probe is covered by a protective sleeve, remove and discard the
protective sleeve.
3. Use a disposable clean soft cloth or paper towel to wipe off the ultrasonic gel or other visible
dirt on the surface of the probe.
4. Choose appropriate cleaners, including spray, disinfectant wipes, neutral detergent, enzymatic
cleaner, and special enzymatic sponge.
5. Immerse the probe completely in water for at least 1 minute. Use a clean soft cloth or soft
sponge to wipe the probe gently, until no dirt are seen on its surface. Do not use a brush, as it
may damage the probe.
Diagnostic Ultrasound System
Service Manual
5 Maintenance

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TE Air and is the answer not in the manual?

Mindray TE Air Specifications

General IconGeneral
BrandMindray
ModelTE Air
CategoryMedical Equipment
LanguageEnglish

Related product manuals