Revision:1.0(2023–5–17)
64
Wireless probe
Severity
Maintenance as soon as possible
Response
Contact the device provider for onsite troubleshooting.
Possible Causes
1. There is severe interference in the environment.
2. The probe is faulty.
Troubleshooting
1. Move to a place with less interference.
2. If the problem persists, the probe is faulty.
Solution
1. Move to a place with less interference.
2. Replace the probe. For details, see the instructions for the i3P ultrasound probe (FDA) 6.1.2
Disassembly and Assembly
.
77..22..22..33 Wrong Password
Fault Description
The input password is wrong.
Involved FRU
Wireless probe
Severity
Maintenance as soon as possible
Response
Handle the error yourself.
Possible Causes
Diagnostic Ultrasound System
Service Manual
7 Troubleshooting