Revision:1.0(2023–5–17)
68
Troubleshooting
Check the self-check result and logs.
Solution
Replace the probe. For details, see the instructions for the i3P ultrasound probe (FDA) 6.1.2
Disassembly and Assembly
.
77..22..33..55 Short Battery Life During Scanning
Fault Description
The battery life is short during scanning.
Involved FRU
Wireless probe
Severity
Maintenance as soon as possible
Response
Contact the device provider for onsite troubleshooting.
Possible Causes
The battery level decreases normally.
Troubleshooting
Check whether the battery self-check result is normal. If so, the battery life results from normal
decrease of the battery level.
Solution
Replace the probe. For details, see the instructions for the i3P ultrasound probe (FDA) 6.1.2
Disassembly and Assembly
.
7.2.4 Charging Failures (Charging Station)
77..22..44..11 Probe Charging Failure
Fault Description
Diagnostic Ultrasound System
Service Manual
7 Troubleshooting