Revision:1.0(2023–5–17)
72
Possible Causes
1. The charging port is not reliably connected.
2. The rating of the charging adapter is low.
3. The ambient temperature is high.
4. The probe is faulty.
Troubleshooting
1. Remove and reconnect the probe and check whether the charging speed is improved.
2. Check whether the rating of the charging adapter is 5 V/3 A.
3. Charge the probe at the room temperature and check whether the charging speed is improved.
4. If the problem persists, the probe is faulty.
Solution
1. Remove and reconnect the probe properly.
2. Replace the adapter with a 5 V/3 A one.
3. Charge the probe at the room temperature.
4. Replace the probe. For details, see the instructions for the i3P ultrasound probe (FDA) 6.1.2
Disassembly and Assembly
.
7.2.6 Charging Failures (Charging Case)
77..22..66..11 Probe Charging Failure
Fault Description
The probe cannot be charged.
Involved FRU
Wireless probe and its Charging Case (Air Capsule)
Severity
Maintenance as soon as possible
Response
Contact the device provider for onsite troubleshooting.
Possible Causes
1. The temperature inside the probe is high.
Diagnostic Ultrasound System
Service Manual
7 Troubleshooting