Revision:1.0(2023–5–17)
62
1.
The battery level is low.
2. The probe is faulty.
Troubleshooting
1. Charge the probe and check whether it can be turned on.
2. If the problem persists, the probe is faulty.
Solution
Replace the probe. For details, see the instructions for the i3P ultrasound probe (FDA) 6.1.2
Disassembly and Assembly
.
77..22..11..22 Shutdown Failure
Fault Description
The probe cannot be turned off.
Involved FRU
Wireless probe
Severity
Maintenance required
Response
Contact the device provider for onsite troubleshooting.
Possible Causes
1. There are bugs in the software.
2. The probe is faulty.
Troubleshooting
1. Hold down the power key of the probe for 6s, and check whether the probe is forcibly turned
off.
2. If the problem persists, the probe is faulty.
3. If the probe is turned off, restart it and check whether the problem is solved.
Solution
Diagnostic Ultrasound System
Service Manual
7 Troubleshooting