Revision:1.0(2023–5–17)
65
The password is forgotten.
Troubleshooting
None
Solution
Restore factory settings of the probe.
7.2.3 Image Errors
77..22..33..11 Abnormal Interruption of Image Display
Fault Description
Display of images is interrupted abnormally.
Involved FRU
Wireless probe
Severity
Maintenance as soon as possible
Response
Contact the device provider for onsite troubleshooting.
Possible Causes
1. The probe is faulty.
2. The Wi-Fi connection is interrupted.
Troubleshooting
1. Replace the mobile phone and try again.
2. Move to a place with less interference.
3. If the problem persists, the probe is faulty.
Solution
1. Replace the mobile phone.
2. Move to a place with less interference.
Diagnostic Ultrasound System
Service Manual
7 Troubleshooting