6. Use a large amount of clean water (about 7.5 L/2 gallon) to wash the probe thoroughly for
about 30s under room temperature, so as to remove residual dirt and cleaning solvent. Repeat
the cleaning operation twice.
7. Use a disposable clean soft cloth or paper towel to wipe the probe.
8. Do not dry the probe by heating it.
9. Check the probe. If there are still visible dirt on the probe, repeat the preceding steps until the
probe is completely cleaned.
Do not immerse the charging cable of the probe in water, disinfectant, or any other
liquid, as doing so may cause electric shock or malfunction.
Diagnostic Ultrasound System
Service Manual
5 Maintenance