3 Device labeling
4 Protection
ground
impedance
Ω Limit: 0.2 Ω
5 Earth leakage
current
Normal condition
(NC)
____μA Limit
NC: 500 μA
SFC: 1000 μA
Single fault
condition (SFC)
____μA
6*
Patient leakage
current
Normal condition
(NC)
□BF____μA Limit
BF applied type
NC: 100 μA
SFC: 500 μA
Single fault
condition (SFC)
□BF____μA
7* Leakage current on the major applied
part
□BF____μA Limit
BF applied type:
5000 μA
8* Patient auxiliary
current
Normal condition
(NC)
□BF____μA Limit
NC: 10μA
SFC: 50μA
Single fault
condition (SFC)
□BF____μA
Name: _____________________________ Date/Signature: