8-95
Document Name Creation
Time
Module Checksum Version Remarks
02_Power
V1.7.pkg
2011-01-20
22:04:00
Module software \ Power
Note
This is an example of the software version/checksum table. View the technical
bulletin to obtain the table that maps the software version being installed.
(5) Select Open.
(6) In the displayed dialog box, check that the creation time, upgrade items, checksum,
version and remarks of each file are correct. Then click OK.
(7) Start the anaesthesia machine and wait for 10 seconds before proceeding with the
next step.
(8) Click Start (standalone) on the Mindray anaesthesia machine software upgrade
tool.
(9) Shut down the anaesthesia machine and then start it again within less than 1 second.
(10) The Windows XP network icon shows a connection success.
Note
If this prompt is not displayed on the PC after the anaesthesia machine is started
normally and performs self-test, check the IP address settings and network cable.
(11) The software update progress is displayed on the screen of the anaesthesia machine.
(12) After the upgrade is completed, Succeed is displayed on the Mindray anaesthesia
machine. A message indicating system update success is displayed on the Mindray
anaesthesia machine software upgrade tool.
(13). Repeat steps 3-12 to upgrade the 02_Powre V1.7.pkg software.
Note
The software upgrade of the power board will usually fail at the first time and will
succeed at the second time.