(14) After the 02_Power V1.7.pkg software is upgraded, shut down the anaesthesia
machine and then start it again.
Note
If the anaesthesia machine cannot be started normally after the upgrade, find the
guide document attached with the upgrade package and take measures according
to the document.
(15) Choose Main Menu > Service to view the software version of the anaesthesia
machine. Input the service password 789789 and press Enter to access system
information and view the software version. Check that the software version of the
anaesthesia machine complies with the table below.
Module Software Version Date*
Host software 03.14.00.07
BIOS 01.15.00.00
FPGA display drive 1.4 \
FPGA sound drive 1.5 \
Auxiliary VCM V01.03.00
VCM V01.04.00
Power supply system V1.7
EFCS software V01.00.00
EFCS-PM software V01.00.00
* The date format may vary according to the equipment settings.
Note
This is an example of the software version table. View the technical bulletin to
obtain the table that maps the software version being installed.