5-42 Image Optimization 
  Description of parameters: 
Method 
Function: select the image acquisition method. 
Selection: Rocked, Linear. 
Linear mode: during the sweep, the probe must be kept parallel. The scanning 
speed should be constant. 
Rocked mode: in this mode, the probe must be moved to a position where you 
can clearly see a middle cut of the object you want to scan and render. Tilt the 
probe to about 30 degrees until the object you want to scan disappears. Start 
the acquisition and tilt the probe over a distance of around 60 degrees until the 
object disappears again. During the sweep, the probe may not be moved 
parallel, just tilted. 
Tip: the speed is related to scanning distance or angle. 
Distance 
Function: to set the distance the probe covered from one end to the other end 
during a linear sweep. 
Range: 10-200 mm, in increments of 10 mm. 
Angle 
Function: to set the angle the probe covered during a fan sweep. 
Range: 10-80°, in increments of 2°. 
5.13.4.3  Smart 3D Image Viewing 
In VR viewing, the system supports the following functions: 
  Render setting. 
  B-mode parameter adjustment. 
  Setting the display format. 
  Viewing MPR. 
  Image zooming. 
  Rotation. 
  Image editing. 
  VR parameter adjustment. 
  Scanning plane and probe movement 
Move the probe across the body surface. 
The arrow in the figure below indicates the movement of the probe (you can move the 
probe in the opposite direction to the arrow.).