Troubleshooting
Error Messages
The following error messages can appear in the Log View frame in the
StakMax Software window.
Table 6-1 Troubleshooting table
Error
Code
Error Message Description Solution
200 Error: A problem was
encountered during
Reset.
Homing routine was
not able to find a
state change on the
home flag,
sensor/flag problem
likely source.
Press the
Power/Reset button.
201 Error: Call Molecular
Devices Technical
Support at +1-800-
635-5577.
Timer resource was
not available,
apparently not used
in code.
Call Molecular Devices
Technical Support +1-
800-635-5577.
202 Error: A problem
occurred during a
motion.
Stall occurred during
motion, encoder and
step counts exceeded
maximum threshold.
Check for object
blocking movement
and then re-align.
203 Error: A problem
occurred with the X
axis.
X Axis motion did not
cause state change
on edge detect switch
—poorly gripped
plate, plate not
present.
Check to make sure
plate is properly
gripped and then re-
align. If error
persists, call
Molecular Devices
Technical Support +1-
800-635-5577.
204 Error: A problem was
encountered during
Reset.
Can't move axis if it
hasn't been homed.
Press the
Power/Reset button.
205 Error: A problem
occurred during a
motion.
Can't move axis if it's
already in motion.
Press the
Power/Reset button.
206 Error: A Problem
occurred with the
EPROM.
Generic EPROM error. Call Molecular Devices
Technical Support +1-
800-635-5577.
207 Error: A problem
occurred during a
motion.
Carriage drop off
exceeds maximum
travel of X Axis.
Re-align the
instrument.