system for leaks. Check the
function of the RO process.
(13) No Pressure,
check pressure sensor!
Check the main water supply.
Check the pressure sensor.
Inform the customer service.
(15) Warning: T3 is
over 40°C!
System temperature caused
by UV lamp or recirculation
pump too high.
Check Interval-Mode. Check the
temperature of the environment
and the feed water.
(16) Tanksensor
overflow! Check
tanklevel sensor!
Product water tank
overflowing.
Check the Tank-settings. Check
the tanklevel sensor.
(17) Warning: Q3 is
over limit!
Replace the polisher module
UPW-Pak.
Reset the expired operating
hours.
(18) OPERATING
HOUR LIMIT ProPak
R10, Exchange
necessary
Pretreatment module
ProPak-R10 exceeded the
maximum operating hours.
Replace the pretreatment module
ProPak-R10. Reset the expired
operating hours.
(32) Please Disinfect
System!
Maximum time until the next
required disinfection has
expired.
Table 5: Display warning messages