membraPure GmbH
Wolfgang-Küntscher-Str.14
16761 Hennigsdorf
Germany
Tel.: +49 (0) 3302/201-20 0
Fax.: +49 (0) 03302/201-20 21
E-Mail: info@membrapure.de
Internet: www.membrapure.de
6 Care and maintenance
6.1 Intervals
Replacement of sterile
filter
Replacement of
pretreatment module
ProPak-R10
Every 2 – 6 months
depending on feed water
quality
OPERATING HOUR LIMIT
ProPak R10
Q1 is too high! (18)
Replacement of
conditioning module
Every 2 - 6 months
depending on feed water
hardness
OPERATING HOUR LIMIT
C.MODUL (3)
EDI ERROR: Q2 too high!
Replacement of polisher
module UPW-Pak
Every 6 – 12 months
recommended
OPERATING HOUR LIMIT UPW
Pak (5)
Warning: Q3 is too high!
Replacement of ultra-
filtration module
Q3 will be incorrect displayed!
Change UFPW! (1)
TOC will be incorrect! Please
change UV-lamp (6)
Replacement of UV-
submersible-lamp
OPERATING HOUR LIMIT UVT
(7)
Replacement of reverse
osmosis pump
OPERATING HOUR LIMIT P1
(2)
Replacement of
circulation pump
OPERATING HOUR LIMIT P2
(4)
Replacement of reverse
osmosis module
Q1 is over limit! Check RO
process (11)
Maintenance by
customer service
Every 12 months
recommended
Table 6: Service intervals
6.2 Replacement of sterile-filter
Depending on feed water quality, ambient conditions and utilization germs are accumulated
in the sterile filter and will contaminate the system.
Install the sterile filter with caution to avoid damage to the filter or the unit.
- Remove the old sterile filter by turning the sterile filter counterclockwise from the
designated port.
- To avoid any leaks, wrap the threads of the sterile filter with Teflon tape.