9.5 Software parameters
9.5.1 Configuration file (config.mcf)
TRUE -> Change all Packs is active
TRUE -> Water production is running
TRUE -> LF1 is in working range
TRUE -> LF2 is in working range
Init value of Tank Sensor
Init value of TOC (abs (Q3-Q4))
dbl_q_tank_volt_limit_warning
(0,01V) on all sensors -> COM failure!
After 2 mins back to main
Refresh memory display (1 min)
Update operating hours (10 mins)
Check P1 counter (30 min)
Counter for operating hours (1 min)
dbl_sec_data_high_density
Acquire data every 1 hour
Acquire data every 24 hours
Acquire data every 12 hours
Init flushing Con.M. (1 hr)(AQ²E only)
dbl_sec_flush_change_cm_edi
Flushing Con.M. after change (1 hr)
Init flushing PropPak (5 min)
Flush Con.M. when tank full (0 sec)
dbl_sec_flush_edi_go_idle
5 mins flush EDI when Tank full
Flush RO when tank full (1 min)
5 mins flush UF after change
dbl_sec_flush_upw_uf_sv6_close
Flushing PW -> SV6 close (15 sec)
dbl_sec_flush_upw_uf_sv6_open
Flushing PW -> SV6 open (15 sec)
dbl_sec_init_flush_upw_no_uf
5 mins first flush UPW-Pack
dbl_sec_init_flush_upw_uf
45 mins first flush UF module
dbl_sec_init_uf_sv6_close
Init flushing PW -> SV6 close (10 sec)
Init flushing PW -> SV6 open (10 sec)
Handshake with control PCB (30 sec)
Switch Messages every (2 sec)
Time to show error NO COMM (2 mins)
PS1 problem is over, 10s wait, back
PS1 problem, close SV1 for 5s
PS1 problem, open SV1 for 2s
Recirculating time (5 min) (when full)
System sleep time (3 hr) (when full)
30 mins Pressure problem -> STOP
15 secs Pressure problem -> Warning