Services FAQ’s
Page 64
9.5.3 Operating hours file (op_hour.dat)
10 Services FAQ’s
Q:
How to change the maximum limit of conductivity sensors?
A:
Max. limit of conductivity Q1 & Q2 can be adjusted in the menu (only for service; see
chapter 9.1.2)
Q:
How to change the maximum limit of temperature T3?
A:
Search for parameter dbl_t3_limit_warning in config.mcf with using .mcf-editor (only for
service; see chapter 9.1.13), change the value, then click “Save Config”
Q:
How to calibrate the touchscreen?
A:
See chapter 9.1.14 for complete guide
dbl_q4_t_temp_min
0
For Calibration!
dbl_tank_limit_high
100
Tank level limit to stop production
dbl_tank_limit_low
70
Tank level limit to start production
dbl_tank_vol_max
100
For Calibration!
dbl_tank_vol_min
0
For Calibration!
dbl_tank_vol_value
-
dbl_tank_volt_max
3,6
For Calibration!
dbl_tank_volt_min
0
For Calibration!
dbl_tank_volt_value
-
Read sensor
dbl_toc_factor
0,1
multiplicator of (Q3-Q4)
dbl_toc_offset
0
offset of (Q3-Q4)
str_serial_number
RW_ _ _ _ _ _
after Endtest, get a serial number
u16_hour_ro_flush_idle_int
12
12hr flush interval when idle
u16_mins_ro_flush
5
Flushing RO (5 mins) when starting
u16_mins_ro_flush_idle
20
Flushing RO (20mins) wenn idle
u16_mins_upw_flush_desinf
50
Disinfection: 50 mins flushing UPW
u16_mins_upw_recirc
10
Disinfection: 10 mins recirculation UPW
u32_uvr_counter
0
how many times UVPW is switched ON
u8_desinf_reminder
2 (6 Months)
Reminder [selector]
Ini-key
Default value
Help
limit_time_cm
850
850 hours
limit_time_p1
8000
8000 hours
limit_time_p2
8000
8000 hours
limit_time_pro_pack
850
850 hours
limit_time_uf_pw
8000
8000 hours
limit_time_upw
8000
8000 hours
limit_time_uvr
8000
8000 hours
limit_time_uvt
8000
8000 hours