6.4 Replacement of the UV-lamp
Figure 16: UV-lamp
- Follow the step in chapter 5.2.2.
- Open the system.
- Put the UV-reactor out of the clips.
- Loosen the UV-lamp by unscrewing the union nut.
- Remove the cap of the UV oxidator.
- Unplug the UV lamp and remove it.
- Insert the new UV-lamp carefully without touching the glass with your fingers.
- Plug the cap on the new UV lamp.
- Fix the UV-lamp by screwing the union nut.
- Put the UV-reaktor into the clips.
- Close the front door.
- Reset the operation hours counter. Please follow the steps in chapter 5.2.2 or
manually in chapter 5.4.3 to do that.
Union nut
cap
UV-lamp
UV-reactor