- Attach the new sterile filter at the dispenser by turning the sterile filter clockwise in
the dedicated port.
- At first water dispensing please open the deaeration screw for a short time to vent
the sterile filter (see Figure 13: sterile filter).
6.3 Modul replacement
Figure 15: Modules in opened system
- Open the system.
- For exchanging of consumables (pretreatment module ProPak-R10, conditioning
module or polisher module UPW-Pak) follow the steps listed in chapter 5.2.1
- For pulling off the tubes from the modules and pluging the tubes onto the new
modules see chapter 4.5, 4.7
- After the flushing and resetting the operating hours counter, the system
automatically jumps into normal operation.
- At first water dispensing please open the deaeration screw for a short time to vent
the sterile filter (see Figure 13: sterile filter).
- Check impermeability on all connections
- Close the front door, the system is ready for operation.
The highest ultra pure water quality is only reachable, if the resin is swollen. This
should be completed within 24 hours after module replacement or first start.
Pretreatment module
ProPak-R10
(black-arrow)
Conditioning module
(red-arrow)
polisher module
UPW-Pak
(blue-arrow